dc.description.abstractWhen capacitive micromachined ultrasonic transducers are immersed in water, the bandwidth of the device is limited by the membrane's second resonance frequency. At this frequency no mechanical power to immersion medium can be transferred. We present a membrane shape to shift the second resonance frequency to a higher value. The structure consists of a very thin membrane at the outer rim with a rigid mass at the center. The stiffness of the central region moves the second resonance to a higher frequency. This membrane configuration is shown to work better in terms of gain and bandwidth as compared to conventional uniform membranes in both transmission and reception. © 2005 IEEE.en_US
dc.language.isoEnglishen_US
dc.source.titleProceedings - IEEE Ultrasonics Symposiumen_US
dc.relation.isversionofhttp://dx.doi.org/10.1109/ULTSYM.2005.1602923en_US
dc.subjectBandwidthen_US
dc.subjectCapacitanceen_US
dc.subjectMicromachiningen_US
dc.subjectNatural frequenciesen_US
dc.subjectResonanceen_US
dc.subjectCapacitive micromachined ultrasonic transducersen_US
dc.subjectNonuniform membranesen_US
dc.subjectSecond resonance frequencyen_US
dc.subjectThin membranesen_US
dc.subjectUltrasonic transducersen_US
dc.titleImproved performance of cMUT with nonuniform membranesen_US
dc.typeConference Paperen_US
dc.departmentDepartment of Electrical and Electronics Engineering
dc.citation.spage597en_US
dc.citation.epage600en_US
dc.citation.volumeNumber1en_US
dc.identifier.doi10.1109/ULTSYM.2005.1602923en_US


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record