Reverse polarized inductive coupling to transmit and receive radiofrequency coil arrays
buir.contributor.author | Atalar, Ergin | |
dc.citation.epage | 456 | en_US |
dc.citation.issueNumber | 2 | en_US |
dc.citation.spage | 446 | en_US |
dc.citation.volumeNumber | 67 | en_US |
dc.contributor.author | Celik, H. | en_US |
dc.contributor.author | Atalar, Ergin | en_US |
dc.date.accessioned | 2016-02-08T09:48:44Z | |
dc.date.available | 2016-02-08T09:48:44Z | |
dc.date.issued | 2011-06-07 | en_US |
dc.department | National Magnetic Resonance Research Center (UMRAM) | en_US |
dc.department | Department of Electrical and Electronics Engineering | en_US |
dc.description.abstract | In this study, the reverse polarization method is implemented using transmit and receive arrays to improve the visibility of the interventional devices. Linearly polarized signal sources-inductively and receptively coupled radiofrequency coils-are used in the experimental setups to demonstrate the ability of the method to separate these sources from a forward polarized anatomy signal. Two different applications of the reverse polarization method are presented here: (a) catheter tracking and (b) fiducial marker visualization, in both of which transmit and receive arrays are used. The performance of the reverse polarization method was further tested with phantom and volunteer studies, and the results proved the feasibility of this method with transmit and receive arrays. | en_US |
dc.description.provenance | Made available in DSpace on 2016-02-08T09:48:44Z (GMT). No. of bitstreams: 1 bilkent-research-paper.pdf: 70227 bytes, checksum: 26e812c6f5156f83f0e77b261a471b5a (MD5) Previous issue date: 2012 | en |
dc.identifier.doi | 10.1002/mrm.23030 | en_US |
dc.identifier.issn | 0740-3194 | |
dc.identifier.uri | http://hdl.handle.net/11693/21611 | |
dc.language.iso | English | en_US |
dc.publisher | Wiley | en_US |
dc.relation.isversionof | http://dx.doi.org/10.1002/mrm.23030 | en_US |
dc.source.title | Magnetic Resonance in Medicine | en_US |
dc.subject | Fiducial marker | en_US |
dc.subject | Inductive and receptive coupling | en_US |
dc.subject | Interventional device tracking | en_US |
dc.subject | Interventional MRI | en_US |
dc.subject | Phased array coils | en_US |
dc.subject | Reverse polarization | en_US |
dc.subject | Transmit array system | en_US |
dc.title | Reverse polarized inductive coupling to transmit and receive radiofrequency coil arrays | en_US |
dc.type | Article | en_US |
Files
Original bundle
1 - 1 of 1
Loading...
- Name:
- Reverse polarized inductive coupling to transmit and receive radiofrequency coil arrays.pdf
- Size:
- 831.49 KB
- Format:
- Adobe Portable Document Format
- Description:
- Full printable version