Piezoelectric power generation using heart motion
Date
Authors
Editor(s)
Advisor
Supervisor
Co-Advisor
Co-Supervisor
Instructor
Source Title
Print ISSN
Electronic ISSN
Publisher
Volume
Issue
Pages
Language
Type
Journal Title
Journal ISSN
Volume Title
Attention Stats
Usage Stats
views
downloads
Series
Abstract
The presence of pacemakers and implantable cardioverter-defibrillators (ICD) is considered historically a contraindication to magnetic resonance (MR) imaging. Main reason behind this contraindication is the current induced on the pacing leads during the MRI examination which may damage the cardiac tissues by heating or the pulse generator of the pacemaker with a reverse current. In this thesis an approach towards the solution of this problem is stated. It has been shown in previous work that replacing the pacing leads with fiber-optic cables minimizes the current induced on the leads. Drawback of this system is the increase in the power consumption of the pacemaker because of the fiber-optic cables and also necessity of an additional pulse generator circuitry near the heart. In this thesis, feasibility of using a piezoelectric power generator for compensating the increased power consumption is investigated. A novel piezoelectric geometry increasing the effective length in a given volume is designed. With this device the resonance frequency of the generator was decreased and the power output for a given volume is increased compared to standard rectangular piezoelectric bimorphs. When connected to a simple heart phantom the novel design produced 2.83 microwatts power whereas the standard rectangular bimorphs produced less than 1 microwatts. Although the output power is increased with the novel design, it was not sufficient to power a pulse generator circuitry that will be used to pace the heart.