dc.description.abstract | In this thesis, methods for optimization of internal magnetic resonance imaging coils have been developed. As a sample optimization endorectal coils were optimized. As the first step ultimate intrinsic signal-to-noise ratio for internal coils was formulated. This value was used as a measure of performance of the internal MRI coils. The related optimization problem was divided into two sub problems: length optimization and cross-sectional optimization. For the length optimization length of a single loop rectangular coil was optimized in order to obtain maximum intrinsic signal to noise ratio at the prostate region. For the cross-sectional optimization designs with different cross-sectional geometries were implemented and tested in saline solution phantom. The designs that gave a significant improvement over the conventional loop coil were favored and the related design ideas were used for the design of an optimized coil. Finally optimized coil was tested in a saline solution phantom model and its performance was calculated. | en_US |
dc.description.provenance | Made available in DSpace on 2016-07-01T11:09:19Z (GMT). No. of bitstreams: 1 0003239.pdf: 1415498 bytes, checksum: 06ad4af85d859706fd3d055abc5cac3e (MD5) Previous issue date: 2006 | en |
dc.description.statementofresponsibility | Eryaman, Yiğitcan | en_US |
dc.format.extent | xv, 90 leaves, illustrations, graphics | en_US |
dc.identifier.itemid | BILKUTUPB101988 | |
dc.identifier.uri | http://hdl.handle.net/11693/29944 | |
dc.language.iso | English | en_US |
dc.rights | info:eu-repo/semantics/openAccess | en_US |
dc.subject | MRI | en_US |
dc.subject | Internal coils | en_US |
dc.subject | Endorectal coil | en_US |
dc.subject | Intrinsic signal to noise ratio | en_US |
dc.subject | Ultimate intrinsic signal-to-noise ratio | en_US |
dc.subject.lcc | WN445 .E78 2006 | en_US |
dc.subject.lcsh | Magnetic resonance imaging. | en_US |
dc.title | Design and implementation of internal MRI coils using ultimate intrinsic | en_US |
dc.type | Thesis | en_US |
thesis.degree.discipline | Electrical and Electronic Engineering | |
thesis.degree.grantor | Bilkent University | |
thesis.degree.level | Master's | |
thesis.degree.name | MS (Master of Science) |