Patient’s body composition can significantly affect RF power deposition in the tissue around DBS implants: ramifications for lead management strategies and MRI field-shaping techniques

buir.contributor.authorKazemivalipour, Ehsan
buir.contributor.orcidKazemivalipour, Ehsan|0000-0003-4221-2397
dc.citation.epage12en_US
dc.citation.issueNumber1en_US
dc.citation.spage1en_US
dc.citation.volumeNumber66en_US
dc.contributor.authorBhusal, B.
dc.contributor.authorKeil, B.
dc.contributor.authorRosenow, J.
dc.contributor.authorKazemivalipour, Ehsan
dc.contributor.authorGolestanirad, L.
dc.date.accessioned2022-02-09T13:34:57Z
dc.date.available2022-02-09T13:34:57Z
dc.date.issued2021-01-13
dc.departmentDepartment of Electrical and Electronics Engineeringen_US
dc.description.abstractPatients with active implants such as deep brain stimulation (DBS) devices have limited access to magnetic resonance imaging (MRI) due to risks associated with RF heating of implants in MRI environment. With an aging population and increased prevalence of neurodegenerative disease, the indication for MRI exams in patients with such implants increases as well. In response to this growing need, many groups have investigated strategies to mitigate RF heating of DBS implants during MRI. These efforts fall into two main categories: MRI field-shaping methods, where the electric field of the MRI RF coil is modified to reduce the interaction with implanted leads, and lead management techniques where surgical modifications in the trajectory reduces the coupling with RF fields. Studies that characterize these techniques, however, have relied either on simulations with homogenous body models, or experiments with box-shaped single-material phantoms. It is well established, however, that the shape and heterogeneity of human body affects the distribution of RF electric fields, which by proxy, alters the heating of an implant inside the body. In this contribution, we applied numerical simulations and phantom experiments to examine the degree to which variations in patient's body composition affects RF power deposition. We then assessed effectiveness of RF-heating mitigation strategies under variant patient body compositions. Our results demonstrated that patient's body composition substantially alters RF power deposition in the tissue around implanted leads. However, both techniques based on MRI field-shaping and DBS lead management performed well under variant body types.en_US
dc.identifier.doi10.1088/1361-6560/abcde9en_US
dc.identifier.eissn1361-6560
dc.identifier.issn0031-9155
dc.identifier.urihttp://hdl.handle.net/11693/77186
dc.language.isoEnglishen_US
dc.publisherInstitute of Physics Publishing Ltd.en_US
dc.relation.isversionofhttps://doi.org/10.1088/1361-6560/abcde9en_US
dc.source.titlePhysics in Medicine & Biologyen_US
dc.subjectBody modelsen_US
dc.subjectMagnetic resonance imaging (MRI)en_US
dc.subjectDeep brain stimulation (DB)Sen_US
dc.subjectFinite element modelingen_US
dc.subjectRF heatingen_US
dc.subjectSafetyen_US
dc.titlePatient’s body composition can significantly affect RF power deposition in the tissue around DBS implants: ramifications for lead management strategies and MRI field-shaping techniquesen_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Patient_s_body_composition_can_significantly_affect_RF_power_deposition_in_the_tissue_around_DBS_implants_ramifications_for_lead_management_strategies_and_MRI_field_shaping_techniques.pdf
Size:
2.86 MB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.69 KB
Format:
Item-specific license agreed upon to submission
Description: