An arbitrary waveform MPI scanner
Author(s)
Date
2022Source Title
International Journal on Magnetic Particle Imaging
Electronic ISSN
2365-9033
Publisher
Infinite Science Publishing
Volume
8
Issue
1Suppl 1
Pages
1 - 4
Language
English
Type
ArticleItem Usage Stats
4
views
views
1
downloads
downloads
Abstract
In magnetic particle imaging (MPI) systems, impedance matching or tuning circuitry has to be employed at a particular operating frequency to handle the reactive power. In this work, we propose a drive coil design with a Rutherford cable winding comprised of 12 Litz wires twisted together to enable arbitrary waveform (AW) characteristics in an MPI scanner. The AW drive coil achieves a 144-fold reduction in inductance and 12-fold reduction in voltage to generate a given drive field amplitude, compared to a standard drive coil with regular Litz wire windings. With imaging experiments, we show that the proposed design can enable imaging in a wide bandwidth, providing flexibility for different functional imaging applications of MPI.