Barnhill, E.Nikolova, M.Arıyürek, CemreDittmann, F.Braun, J.Sack, I.2020-02-042020-02-042019-070278-0062http://hdl.handle.net/11693/53057MRI phase contrast imaging methods that assemble slice-wise acquisitions into volumes can contain interslice phase discontinuities (IPDs) over the course of the scan from sources, including unavoidable physiological activity. In magnetic resonance elastography (MRE), this can alter wavelength and tissue stiffness estimates, invalidating the analysis. We first model this behavior as jitter along the z-axis of the phase of 3D complex-valued wave volumes. A two-step image processing pipeline is then proposed that removes IPDs. First, constant slicewise phase shift is removed with a novel, non-convex dejittering algorithm. Then, regional physiological noise artifacts are removed with novel filtering of 3D wavelet coefficients. Calibration of two pipeline coefficients, the dejitter parameter α and the wavelet band high-pass coefficient ωc , was first performed on a finite-element method brain phantom. A comparative investigation was then performed, on a cohort of 48 brain acquisitions, of four approaches to IPDs: 1) the proposed method; 2) a “control” condition of neglect of IPDs; 3) an anisotropic wavelet-based method; and 4) a method of in-plane (2D) processing. The present method showed medians of |G∗|=1873 Pa for a multifrequency wave inversion centered at 40 Hz which was within 6% of methods 3) and 4), while neglect produced |G∗| estimates a mean of 17% lower. The proposed method reduced the value range of the cohort against methods 3) and 4) by 29% and 31%, respectively. Such reduction in variance enhances the ability of brain MRE to predict subtler physiological changes. Our theoretical approach further enables more powerful applications of fundamental findings in noise and denoising to MRE.EnglishMagnetic resonance imagingElastographyImage denoisingFast robust dejitter and interslice discontinuity removal in MRI phase acquisitions: application to magnetic resonance elastographyArticle10.1109/TMI.2019.28933691558-254X