Karaman, Mustafa2016-01-082016-01-081992http://hdl.handle.net/11693/18456Ankara : Department of Electrical and Electronics Engineering and Institute of Engineering and Science, Bilkent Univ., 1992.Thesis (Ph. D.) -- Bilkent University, 1992.Includes bibliographical references.For phased array medical ultrasound imaging, digital beamforming techniques and VLSI circuits are studied to form a fully digital receive front-end hardware. To increase the timing accuracy in beamforming, a computationally efficient upsampling schme is examined. A digital beamforming method based on receive focusing at the raster focal points is proposed. The sector images of the resolution phantom, reconstructed from the phased array experimental data by beamforming at the radial and raster focal points, are presented for comparison of the image resolution performance of the two beamforming schemes. For adaptive beamforming, a phase aberration correction technique with very low computational complexity is described. Image quality performance of the technique is examined by digitally processing the non-aberrated and aberrated phased array experimental data sets of an ultrasound resolution phantom. VLSI circuits and their implementations for the proposed techniques are presented.x, 90 leavesEnglishinfo:eu-repo/semantics/openAccessMedical ultrasound imagingDigital beamformingPhased array beamformingPhase aberration correctionVLSI circuitQT34 .K37 1992Ultrasonic imaging.Ultrasonics--Diagnostic use.Integrated circuits.Electron beams.Digital beamforming techniques and VLSI circuits for medical ultrasound imagingThesis