Eryaman, Yiğitcan2016-01-082016-01-082011http://hdl.handle.net/11693/15257Ankara : The Department of Electrical and Electronics Engineering and the Institute of Engineering and Sciences of Bilkent University, 2011.Thesis (Ph. D.) -- Bilkent University, 2011.Includes bibliographical references leaves 86-88.In this thesis, novel methods are presented, which can be used to reduce the heating of the human body due to radiofrequency fields in magnetic resonance imaging (MRI). The proposed methods depend on the modification of the electric field distribution for reducing the specific absorption rate (SAR). These methods can be used to reduce the local SAR in the vicinity of metallic devices and the whole-volume average SAR, as shown by electromagnetic field simulations and phantom, animal and patient experiments. These results can improve the safety of MRI scans performed on patients with metallic implants and MRI-guided interventional procedures. Additionally, by reducing the whole body average SAR, safer and faster MRI scans can be performed.xiv, 88 leaves, illustrations, graphicsEnglishinfo:eu-repo/semantics/openAccessWN185 .E79 2011Magnetic resonance imaging--Methods.Novel SAR reduction methods for magnetic resonance imagingThesis